The inconsistency between the measured temperature and the actual temperature of the laboratory rotary tube furnace may be caused by various reasons, including:
1. Temperature sensor related issues
Improper sensor position
Reason: If the temperature sensor is not placed in a position that can accurately reflect the average temperature inside the furnace, such as near heating elements or furnace doors where there is a large temperature gradient, it will result in a deviation between the measured temperature and the actual temperature.
Example: The sensor is closely attached to the heating zone of the furnace tube, and the measured temperature is too high; If placed at the cold end of the furnace tube, the measured temperature will be too low.
Sensor aging or damage
Reason: Long term high temperature environment may cause sensor performance degradation or measurement distortion due to physical damage.
Testing method: Use a standard thermometer for comparative measurement, or calibrate the sensor regularly.
Sensor response delay
Reason: The sensor’s response speed to temperature changes is insufficient, resulting in a delay in dynamic measurement compared to actual temperature changes.
Impact: During rapid heating or cooling, there is a significant difference between the measured value and the actual value.
2. Thermocouple or compensation wire issues
Thermocouple material mismatch
Reason: The thermocouple material does not match the compensation wire material, resulting in inaccurate cold end compensation.
Impact: Systematic deviation in temperature measurement.
Compensation wire fault
Reason: Compensation wire insulation damage, poor contact, or electromagnetic interference can cause signal transmission distortion.
Testing method: Check the appearance of the wire, measure the resistance value, and eliminate electromagnetic interference sources.
3. Temperature control system issue
Improper PID parameter settings
Reason: PID controller parameters (proportional, integral, derivative) were not optimized, resulting in unstable or overshoot temperature control.
Impact: The measured temperature fluctuates greatly and deviates significantly from the set value.
Calibration deviation of temperature controller
Reason: The temperature controller has not been regularly calibrated or the calibration method is incorrect, resulting in the accumulation of measurement errors.
Solution: Regularly calibrate the temperature controller using a standard temperature source.
4. Furnace structure and insulation performance issues
Poor sealing of furnace body
Reason: The furnace door, flange and other parts are not sealed tightly, resulting in heat loss and actual temperature lower than the measured value.
Detection method: Check the leakage point of the furnace body and repair the sealing components.
Aging of insulation materials
Reason: The performance of insulation materials has decreased, resulting in uneven temperature distribution inside the furnace and deviation between measured values and actual values.
Solution: Replace the aging insulation material.
5. Operation and process setting issues
Heating rate set too fast
Reason: The heating rate exceeds the equipment’s capacity, causing temperature control to lag and the measured value to be out of sync with the actual value.
Solution: Reasonably set the heating rate to avoid temperature overshoot.
The influence of samples on temperature
Reason: The heat absorption or release characteristics of the sample cause local temperature changes, which affect the measured values.
Solution: Optimize the placement of samples or use multi-point measurement of average temperature.
6. Environment and interference factors
electromagnetic interference
Reason: The strong electromagnetic field nearby interferes with the temperature sensor signal, resulting in measurement errors.
Solution: Block the sensor signal line and keep it away from interference sources.
Environmental temperature changes
Reason: The temperature fluctuation in the laboratory environment affects the heat dissipation of the furnace body, resulting in a deviation between the actual temperature and the measured value.
Solution: Control the laboratory environment temperature or insulate the furnace body.
7. Equipment aging and maintenance issues
Long term use of equipment
Reason: Aging of furnace body, heating elements and other components leads to a decrease in temperature control performance.
Solution: Regularly maintain the equipment and replace aging components.
Improper maintenance
Reason: Failure to clean, calibrate, and maintain the equipment according to regulations resulted in inaccurate temperature measurement.
Solution: Establish an equipment maintenance plan and strictly implement it.