Website navigation

TECHNICAL

Position:Home > Technical

Customized experimental muffle furnace daily maintenance

Time:2025-07-22 Click:0
  

As a high-temperature heat treatment equipment, the performance stability of the experimental muffle furnace directly affects the accuracy of the experimental results. To extend equipment lifespan and reduce failure rates, it is necessary to develop a daily maintenance plan for the system. The following is an explanation from five aspects: cleaning, inspection, calibration, operating standards, and environmental maintenance:

1. Daily cleaning and maintenance
Internal cleaning of furnace
Frequency: After each experiment or once a week (depending on usage frequency).
method:
After the furnace cools to room temperature, use a soft bristled brush or vacuum cleaner to remove residual oxides and sample debris.
Stubborn stains can be wiped with anhydrous ethanol or specialized cleaning agents, avoiding the use of corrosive solvents.
Regularly check whether the drainage holes at the bottom of the furnace are blocked to ensure that the condensate can be discharged normally.
Attention: Wear heat-resistant gloves during cleaning to avoid scratching the inner wall of the furnace (such as alumina ceramic fiber board).
Maintenance of furnace doors and seals
Frequency: Once a month.
method:
Wipe the inside and outside of the furnace door with a soft cloth to remove accumulated dust or spilled samples.
Check whether the sealing strip (such as silicone or asbestos rope) is aging or cracking, and replace it in a timely manner if it is deformed.
Apply a small amount of high-temperature resistant grease (such as molybdenum disulfide) on the surface of the sealing strip to enhance its sealing performance.
Ventilation system cleaning
Frequency: Once every quarter.
method:
After turning off the power, open the back cover or side panel of the device and clean the filter or filter cotton of the air inlet and exhaust outlet.
Check whether there is dust accumulation on the blades of the circulating fan, blow them clean with compressed air or clean them with a soft bristled brush.
Ensure that the ventilation ducts are not bent or blocked, and that the airflow is smooth.

2. Key component inspection
Heating element inspection
Frequency: Once every six months.
method:
Visually inspect whether the silicon carbide rod and resistance wire are broken, oxidized, or deformed.
Measure the resistance value of the heating element with a multimeter (the cold resistance of the silicon carbide rod is usually 5-10 Ω), and replace it if the deviation exceeds ± 10%.
Check if the connecting wires of the heating element are loose or burnt, tighten the screws and apply conductive paste.
Thermocouple calibration and inspection
Frequency: Once a year (or adjusted according to experimental accuracy requirements).
method:
Compare the thermocouple with a standard thermometer (such as a platinum resistance thermometer) and calibrate it at commonly used temperature points (such as 500 ℃, 800 ℃).
Check if the thermocouple protective sleeve is damaged, if the wiring terminals are oxidized, and replace them if necessary.
Ensure that the insertion depth of the thermocouple meets the requirements (usually 1/3-1/2 of the furnace height).
Electrical system inspection
Frequency: Once every quarter.
method:
Check if the power and control cables are damaged, and if the plugs and sockets are loose.
Use an insulation resistance meter to measure the insulation resistance of the equipment to ground (should be ≥ 1M Ω) to avoid the risk of electrical leakage.
Check whether the control components such as solid-state relays and contactors are overheating or making abnormal noises, and replace them if necessary.

3. Temperature control system calibration
pid parameter optimization
Frequency: Once a year or after equipment modification.
method:
Record the deviation between the actual temperature curve and the set value through an empty furnace heating test.
Adjust PID parameters based on deviations (such as proportional band P, integration time Ti, differentiation time Td) to control temperature fluctuations within ± 2 ℃.
Refer to the equipment manual or contact the manufacturer to obtain the initial PID parameters to avoid arbitrary modifications that may cause system instability.
Program functional testing
Frequency: Once every six months.
method:
Run the preset program (such as heating insulation cooling) and check if each stage is executed according to the settings.
Test whether the functions of over temperature alarm, power-off protection, etc. are normal to ensure the reliability of the safety interlock device.

4. Operating standards and precautions
Temperature control
To avoid short-term significant temperature rise (such as directly increasing from room temperature to 1000 ℃), it is recommended to set the temperature in sections (such as 200 ℃ → 500 ℃ → 800 ℃ → 1000 ℃).
When cooling down, natural cooling is preferred. If forced cooling is required, ensure that the furnace temperature is below 200 ℃ before opening the furnace door.
Sample placement requirements
The sample should be placed in the central area of the furnace, avoiding proximity to heating elements or furnace walls.
Use high-temperature resistant crucibles (such as alumina and corundum materials), and do not place volatile or corrosive samples.
The total weight of the sample shall not exceed the rated load of the equipment (usually 1/3-1/2 of the furnace volume).
Emergency situation handling
If an over temperature alarm occurs, immediately cut off the power and check the temperature control system after the furnace cools down.
If the furnace door cannot be closed properly, stop using the equipment and check the seals or mechanical structure.
Regularly organize operator training to familiarize themselves with equipment safety operating procedures.

5. Environment and storage conditions
Environmental requirements for use
Temperature: 0-40 ℃, humidity ≤ 85% RH, avoid using in humid or corrosive gas environments.
Ventilation: Reserve at least 30cm of space around the equipment to ensure good heat dissipation.
Power supply: Stable three-phase power (380V) or single-phase power (220V), equipped with independent leakage protection switch.
Long term shutdown maintenance
Clean the furnace and apply rust proof oil, disconnect the power supply and cover it with a dust cover.
Power on once a month (raise the temperature of the furnace to 200 ℃ and maintain it for 1 hour) to prevent electrical components from getting damp.

6. Maintenance records and archive management
Establish an equipment maintenance ledger to record the time, content, and results of cleaning, inspection, calibration, maintenance, and other operations.
Save replacement records of key components (such as heating elements and thermocouples) for easy traceability of equipment historical status.
Regularly analyze maintenance data, predict component lifespan, and develop replacement plans in advance.
Through systematic daily maintenance, the failure rate of experimental muffle furnaces can be significantly reduced, and their service life can be extended to more than 10 years, while ensuring the accuracy and repeatability of experimental data. It is recommended to develop a personalized maintenance plan based on the equipment manual and have it regularly executed by professionals.

Contact Information

E-mail:
web@kejiafurnace.com
Address:
No.18 Hongye Road, Hi-tech zone , Zhengzhou, China, 450000
QQ tel Email WhatApp FaceBook

Tel Number

181-3719-5600

Email

web@kejiafurnace.com

WhatsApp

+86/18037178440

FaceBook

Zhengzhou/Kejia

kejialiuyan